Pengaruh Desain Modular Terhadap Stabilitas Suhu dan Reduksi Kebisingan pada Incu Analyzer 452 296

Authors

DOI:

https://doi.org/10.26623/transformatika.v23i2.12330

Keywords:

Incu Analyzer, Inkubator Bayi, Sensor Suhu, Sensor Tingkat Kebisingan, Kesalahan Nilai

Abstract

Incu analyzer is a tool used to measure temperature, airflow noise and humidity. The way the tool works during calibration is placed on a mattress. The temperature sensor used is an LM35 sensor that produces a linear voltage value with temperature. The output voltage is proportional to the temperature in degrees Celsius. The noise level sensor uses a sound sensor microphone module, which is an electronic module that functions to detect sound and convert it into an electrical signal. This Sound Mic Microphone Sensor Detection Module for Arduino has a sound detection range of ± 30 cm to 1 meter. When testing the 35°C setting, the error percentage was 7%, the 36°C setting obtained a 3% error percentage and the 37°C setting obtained a 3% error percentage. The noise level sensor test data was carried out 9 times at 5 minute intervals. And the average error value was 1.2 (dB), and the overall error value was 0.92 (dB).

 

Downloads

Download data is not yet available.

References

Y. S. Nafie, J. Tarigan, A. Ch Louk, and K. Kunci, RANCANG BANGUN SISTEM KONTROL PARAMETER FISIK PADA INKUBATOR BAYI BERBASIS MIKROKONTROLER ARDUINO UNO DAN ESP 8266.

E. Arianto and A. Siswoyo, Analisis Tingkat Penyebaran Suhu Inkubator Bayi Dengan Sensor DS18B20, J-Innovation, vol. 11, no. 2, pp. 39–43, Dec. 2022, doi: 10.55600/jipwa.v11i2.144.

A. Taqwa et al., Monitoring Temperature Bayi Dengan Sistem Wireless Sensor Network Berbasis Arduino Uno ATmega328.

Published

2025-01-14 — Updated on 2026-01-13

How to Cite

prasojo, fajri, Andi Kurniawan Nugroho, & Sri Heranurweni. (2026). Pengaruh Desain Modular Terhadap Stabilitas Suhu dan Reduksi Kebisingan pada Incu Analyzer. Jurnal Transformatika, 23(2), 155-163. https://doi.org/10.26623/transformatika.v23i2.12330